Description
Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.
Problem types
CWE-400 Uncontrolled Resource Consumption
Product status
Any version
VG2.3.2 (custom)
References
static.draeger.com/...VGxx-Security-advisory-update_v1-3.pdf
www.vulncheck.com/...300-vg2-network-based-denial-of-service