Home

Description

The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.

PUBLISHED Reserved 2024-10-10 | Published 2024-11-14 | Updated 2024-11-18 | Assigner Baxter




CRITICAL: 9.3CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

Problem types

CWE-494 Download of Code Without Integrity Check

Product status

Default status
unaffected

06.08.00.00 and prior
affected

References

www.cisa.gov/...vents/ics-medical-advisories/icsma-24-319-01

cve.org (CVE-2024-48974)

nvd.nist.gov (CVE-2024-48974)

Download JSON